[CTK Biotech] CK-MB FIA Test
CK-MB
REAGENTS AND MATERIALS PROVIDED:
- Individually sealed foil pouches containing:
a. One test device
b. One desiccant - Detection buffer tubes
- SD card
- Instructions for Use
MATERIALS REQUIRED BUT NOT PROVIDED:
- Clock or timer
- RaFIA Immunofluorescence Analyzer
- RaFIA Immunofluorescence Incubator
- Quantitative, non-subjective results
- Increased sensitivity and broader detection range compared to rapid tests
RANGES
- Working range: 1.0-80.0 ng/mL
PRECISION
- Intra-lot Precision
Intra-lot precision was determined by testing of CK-MB reference materials using 10 test devices from the same lot. CV ≤ 15%. - Inter-lot Precision
Inter-lot precision was determined by testing of CK-MB reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 15%.
ACCURACY
- CK-MB control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
LINEARITY
- A serial concentration of CK-MB reference materials at 2.5-80.0 ng/mL was tested, and the correlation coefficient (R) is ≥0.9900.